OLV-8000C Multi-Parameter Patient Monitor

OLV-8000C Multi-Parameter Patient Monitor

AU Plug
Sale price  $999.00 USD Regular price  $1,500.00 USD
Skip to product information
OLV-8000C Multi-Parameter Patient Monitor

OLV-8000C Multi-Parameter Patient Monitor

Sale price  $999.00 USD Regular price  $1,500.00 USD
Plug

Advanced, expandable, and intuitively designed for critical care environments. The Resoxy OLV-8000C is a full-featured patient monitor built around a high-resolution 12.1-inch color TFT touch screen. It delivers precise, real‑time tracking of ECG, NIBP, SpO₂, RESP, and dual‑channel TEMP—with optional expansion to EtCO₂, IBP, and central monitoring. Engineered for durability and clarity, it supports adult, pediatric, and neonatal care across the ICU, OR, emergency transport, and general wards.

Key Features

  • 12.1-Inch High-Resolution Touch Screen: Vivid real‑color display for clear waveform visualization, ST‑segment analysis, pacemaker detection, and intuitive menu navigation.

  • Comprehensive Multi-Parameter Monitoring: Standard configuration includes ECG (3/5‑lead), NIBP, SpO₂, RESP, and 2‑channel TEMP. Optional modules add EtCO₂, IBP (ART/PA/CVP), trolley mounting, and wireless central monitoring.

  • Clinical-Grade Precision: Features diagnostic‑grade ECG bandwidth (0.05‑130Hz), high CMRR (>100dB), accurate ST‑analysis (±0.02mV), and validated NIBP/SpO₂ measurements with 3‑level visual/audio alarms.

  • Robust & Portable Design: Weighs only 4.5kg with a built‑in rechargeable lithium battery for transport between departments. IP‑rated housing, defibrillator‑proof circuitry, and a 5‑year operational lifetime.

  • Smart Data Management: Stores 400 NIBP readings, 6000‑second ECG waveform recall, 60 alarm events, and 7‑day trend charts. Supports wired/wireless central monitoring networks.

Technical Specifications

  • Display: 12.1” Real Color TFT Touch Screen

  • Standard Parameters: ECG, RESP, NIBP, SpO₂, PR/HR, 2‑TEMP

  • Optional Parameters: EtCO₂, IBP (ART/PA/CVP/ICP)

  • Power: 220VAC, 50Hz | Built‑in 11.1V 2200mAh Li‑ion Battery

  • Connectivity: Network port for central monitoring, USB

  • Size: 342 × 252 × 139.5 mm | Weight: 4.5 kg

  • Compliance: CE Certified, Class I Equipment

  • Operating Environment: 0‑40°C, 15‑80% RH

Monitor Includes
Resoxy OLV‑8000C Monitor, Power Cable, ECG Cable & Electrodes, NIBP Cuff & Hose, SpO₂ Sensor, Temperature Probes, User Manual.

Expandable. Precise. Always Connected.
The OLV‑8000C combines hospital‑level accuracy with the flexibility needed for modern dynamic care—ensuring every vital parameter is monitored with clarity and confidence.

Note: Optional parameters and accessories (trolley, bracket, EtCO₂/IBP modules) available separately.


Standard Configuration

SpO2, PR/HR, NIBP, ECG, RESP, 2-TEMP

✔12.1 inch high-resolution color TFT LCD screen

✔Thinner and lighter design, convenient and portable for user

✔ Freeze instant waveform for real time S-T segment analysis, pace-maker detection.

✔ Full touch screen selectable, more convenient for operation

✔ Real time S-T segment analysis, pace-maker detection

✔ 3-level audio/visual alarming

✔ Efficient resistance to interference of defibrillator and HF knife.

✔ Lasting working capacity with built-in rechargeable lithium battery

✔ 400 groups of NIBP list, 6000 seconds ECG waveform recall, 60 alarm evens records recall, 7-day trend chart in storage

✔ Wire/wireless central monitoring system applicable to ICU/CCU/OR

▲ 1 ECG Specification
1.1 Lead configuration :Standard 3-lead or 5-lead
3-lead :RA, LA, LL,Lead method:I,II,III
5-lead :RA, LA, LL, RL, V,Lead method:I,II,III,aVR,aVL,aVF,V
1.2 Gain variation :0.25, .5, 1, 2. Automatically,Gain change per minute < 0.66%/min,Total gain change in 1 hour < ±10%
1.3 HR Range :
Adult :15 ~ 300bpm (beat/minute)
Precision : ±1% or ±1bpm,the larger prevails
Resolution :1 bpm(beat/minute)
1.4 Sensitivity :> 200 uV (Peak-to-peak value)
1.5 Input Impedance :> 5 (megohm)
1.6 Bandwidth
Diagnostic mode :0.05~130Hz
Monitoring Mode :0.5~40Hz
Operation mode :1~20Hz
1.7 Common Mode rejection Ratio
Diagnostic mode > 90 dB
Monitoring Mode > 100 dB
Operation mode > 100 dB
1.8 Pole Polarization Vvoltage Range :±300mV
1.9 Pacing Pulse Test
Test pacing pulse in accordance with the following conditions :
Amplitude : ±2 mV ~ ±700mV
Width : 0.1ms ~ 2ms
Risetime : 10us~100µs
1.10 Pacing pulse inhibition
When pacing analysis switch is on, pacing pulse in accordance with the following conditions are
restrained, but affection against HR calculation.
Amplitude : ±2 mV ~ ±700mV
Width : 0.1ms ~ 2ms
Risetime : 10us ~ 100µs
1.11 Signal Range :±8 mV (Peak-to-peak value)
1.12 Calibrating Signal :1mV(Peak-to-peak value),precision ±5%
1.13 ST Segment Measuring Volume
Measuring range : -2.0mV ~ +2.0mV
Measuring precision : Ranging -0.8 mV ~ +0.8mV, measuring error is ±0.02mV or ±10%,. The larger prevails. No definition for other ranges.
▲ 2 RESP Specification
2.1 Measuring Method :RA-LL impedance
2.2 RESP Impedance Measuring Range :0.3~3Ω
2.3 Base Impedance Range :200~4000Ω
2.4 Bandwidth :0.1~2.5Hz
2.5 RESP Rate
Range
Adult :0~120BrPM
Resolution :1 BrPM
Precision :±±2 BrPM
2.6 Asphyxia Alarm :10~40 seconds
▲ 3 NIBP Specification
3.1 Measuring Method :Pulse wave oscillometry
3.2 Work Mode :Manual/Automatic/STAT
3.3 Measuring Interval of Automatic Measuring Mode :1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
3.4 Measuring Time of STAT Mode :5 minutes
3.5 PR range :40 – 240 bpm
3.6 Measuring Range and Precision
Range
Adult :Systolic blood pressure 40~270mmHg
Diastolic blood pressure 10~215mmHg
Mean blood pressure 20~235mmHg
Static pressure range : 0~300mmHg
Static pressure precision : ±3mmHg
Pressure precision : Max. average error: 5mmHg;Max. standard deviation: 8mmHg
3.7 Overvoltage protection
Adult mode :300 mmHg
▲ 4 SpO2 Specification
4.1 SpO2
Range : 0~100%
Resolution : 1%
Precision : 70~100%:±2%
0%~69% :no definition given
4.2 PR
Range 20~300bpm
Resolution 1bpm
Precision ±±3bpm
▲ 5 TEMP Specification
5.1 suitable for temperature sensor :YSI series,CYF series
5.2 Channel quantity :2 channels
5.3 Measuring
Range :0~50℃
Resolution :0.1℃
Precision : ±0.1℃(excluding sensor error)
▲ 6 IBP Specification
Label :ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART :0 ~ 300 mmHg
PA :-6 ~ 120 mmHg
CVP/RAP/LAP/ICP :-10 ~ 40 mmHg
P1/P2 :-10 ~ 300 mmHg
Press Sensor
Sensitivit : 5 uV/V/mmHg
Impedance : 300-3000Ω
Resolution :1 mmHg
Accuracy :±2% or ±1mmHg, which great
Actualization interval :about 1 Sec.
▲ 7 CO2 Specification
Side stream :
7.1 measurement :Infrared radiation absorption technique
7.2 measurement range :0 ~ 99mmHg
7.3 precision* :0 ~ 40mmHg : ±2mmHg
41 ~76mmHg : ±5%
77~99mmHg : ±10%
7.4 resolution :1mmHg
7.5 drift :To meet the requirement of accuracy within 6 hours
7.6 pumping rate :70ml/min, 100ml/min
7.7 pumping rate control accuracy :15% or 15ml/min big
7.8 module start time
The < 1min, enter the precision measurement of the state,After entering 1min, the accuracy of measuring state
7.9 AwRR measuring range :0 ~120BrPM: ±2
7.10 precision
0 ~70BrPM : ±2BrPM
>70BrPM : ±5BrPM
7.11 response time
The use of adult sampling tube :
< @ 5.5s 100ml/min
< @ 7s 70ml/min
7.12 delay time
The use of adult sampling tube :
< @ 5s 100ml/min
< @ 6.5s 70ml/min
7.13 asphyxia alarm delay
AwRR : 10 ~ 40s










You may also like